Unpacking and visual inspection

Mechanical function test

Cable winch

P
Pull out the cable until it is completely unwound.
Hold the cable with one hand near the duct to slow it down as it is being wound. Pull up
the brake of the cable winch and wind up the cable completely.
This device is equipped with a standard safety plug. Attach a power plug that is compliant
with local standards if necessary.

Support arm movements

P
Unlock the support arm. Then check its movements, including several stops. If the DAP
chamber option is included in the delivery, it must be attached to the collimator. The arm
should be easy to move through its full range using one hand and should stop in every
position without any additional movement (up or down).
Fig. 11: Positioning of support arm

Single tank and collimator movements

P
Turn the collimator with the handle to the end positions ±90°.
Rotate the tube to all of its commonly used working and park positions.