3 - 8
All Mammomats: Removing the cable harness from the left side
of the stand
NOTICE
All connectors on D801,
except flat cable to D802
Remove R805
completely
*1 X885 Remove
X801
Fig. 9
*1 MAMMOMAT 3000 CE: Connectors X885 and X880 are omitted
MAMMOMAT 300 / 3000
for MAMMOMAT 300/ MAMMOMAT 300 CE:
Some of the connectors below are omitted
D801
D802
Register 5
RXB7-230.091.05
Rev. 03
Removing de cable harness
Remove protective conductors
All connectors on D802
Remove all protective conductors
Remove these wires (facing
X880
right) only *1
4
5
6
7
8
green / yellow
X881 Remove all wires
Remove relay socket K801 by removing
the rivets
X886 Remove
Page 8 of 16
07. 98
TD SD 24
blue and yellow
orange and red
red (heavy wire)
Siemens AG
Medical Engineering
3