Chapter G - TROUBLESHOOTING

8.3 Z-carriage gets stuck

Z-carriage gets stuck

8.4 Mechanism does not observe limits

Mechanism does not observe
either of the limits
Mechanism does not observe
one of the limits

9 PROBLEMS IN RADIOGRAPHS

9.1 Disturbances in radiographs

Most often troubles are caused by other reasons, not by the equipment. Check that the fault
on the radiograph does not result from defective patient positioning or fault in the film.

9.2 No radiograph

Exposure light and signal do
not function. If error code
appears, see error code expla-
nations.
G-10 Planmeca Proline EC X-ray unit
SYMPTOM
SYMPTOM
SYMPTOM
POSSIBLE REASON
One of the steel cables broken.
Creeping speed adjusted incor-
rectly.
Cables or cable wheel dam-
aged.
POSSIBLE REASON
Hall sensor does not function.
One of the impulse magnets
too far from the Hall sensor,
installed to a wrong direction or
completely missing.
POSSIBLE REASON
The equipment is in test mode.
PROBLEMS IN RADIOGRAPHS
CORRECTIVE ACTION
CORRECTIVE ACTION
CORRECTIVE ACTION
Technical Manual