Disturbances In Rotation; Rotation Part Does Not Move At All; Stop In The Middle Of The Rotation - Planmeca Proline EC pan Technical Manual

Table of Contents

Advertisement

DISTURBANCES IN ROTATION

4 DISTURBANCES IN ROTATION

4.1 Rotation part does not move at all

Motor gets no guidance
The sound of the motor is audi-
ble

4.2 Stop in the middle of the rotation

The Motor stops. If error code
appears, see error code expla-
nations.
Driving wheel stops (motor
stalls). If error code appears,
see error code explanations.
Technical Manual
SYMPTOM
SYMPTOM
POSSIBLE REASON
Keyboard out of order
Generator processor PCB out
of order
Rotation motor out of order or
connector loose
Transportation
pins
removed
If the rotation part can be
moved with hands, motor or
generator processor PCB faulty
or one motor cable loose
POSSIBLE REASON
No error code: Motor or Gener-
ator processor PCB defective
or one cable loose.
Adjustment of guidance wheel
erroneous.
The equipment is installed atilt. Align the unit properly.
Gear assembly damaged.
One of the step motor connec-
tions defective.
Cable guide stucks.
Chapter G - TROUBLESHOOTING
CORRECTIVE ACTION
Replace the keyboard assem-
bly
Replace the PCB
Check the connections and/or
replace the gear/motor assem-
bly
not
Remove the pins
Check the connections and try
with new generator processor
PCB
CORRECTIVE ACTION
Check the connections and/or
replace the Generator proces-
sor PCB or replace the motor.
Adjust the guidance wheel.
Replace the gear assembly.
Check the connections.
Planmeca Proline EC X-ray unit G-3