Planmeca Proline EC pan Technical Manual page 155

Table of Contents

Advertisement

CEPHALOSTAT ADJUSTMENT
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Adjust the cephalostat head support position to horizontal position, if needed (see Fig. 111, 4
and Fig. 112 below)
Figure 112
Fasten the loosened screws and replace the covers.
NOTE
CHECK THE RADIATION BEAM ALIGNMENT ACCORDING TO THE INSTRUCTIONS
GIVEN IN SECTION "Checking the cephalometric beam alignment" on page E-48
AFTER ADJUSTING THE HEAD SUPPORT POSITION. ADJUST THE BEAM, IF NEEDED.
REPEAT THE BEAM AND HEAD SUPPORT ADJUSTMENTS UNTIL THE RADIATION
BEAM POSITION AND HEAD SUPPORT POSITION ARE ACCEPTABLE.
Technical Manual
Planmeca Proline EC X-ray unit E-61