Operational Instructions and Function Adjustment
Mode
Select the display mode to be adjusted. The following table lists the initial values of all modes.
Mode
Custom
DICOM200
DICOM250
DICOM300
DICOM350
DICOM400
DICOM450
DICOM500
DICOM600
DICOM700
DICOM User
Note
After you select a mode and adjust its value, the LCD monitor displays content based on the
adjusted value.
Brightness
Set Brightness to adjust the monitor brightness. In DICOM200, DICOM250, DICOM300,
DICOM350, DICOM400, DICOM450, DICOM5, DICOM600, or DICOM700 mode, the system
automatically switches to DICOM User mode after brightness is adjusted.
Contrast
Set Contrast to adjust the contrast ratio in the current display mode. The value range is 0 to 100.
The default value is 50.
Note
This item can be adjusted only in Custom mode.
Gamma
Set Gamma to adjust the Gamma curve displayed by an image. The options include DICOM, CT,
MRI, DSA, Gamma 1.8, Gamma 2.0, Gamma 2.2, Gamma 2.4, and Gamma 2.6. The default
value is Gamma 2.2.
Note
This item can be adjusted only in Custom mode.
2
Brightness (cd/m
)
400
200
250
300
350
400
450
500
600
700
400
15
Gamma Value
Gamma 2.2
DICOM
DICOM
DICOM
DICOM
DICOM
DICOM
DICOM
DICOM
DICOM
DICOM