Commissioning
When the calibration process is canceled and the device is reset to
the factory settings, the default values are loaded!
Datalogging
Menu item
This menu item allows you to activate/deactivate the data recording function.
Datalogging inactive
➜
Datalogging active
➜
➜
Confirm your entry with
For detailed information about the data recording function, see
section "Data Recording" on page 120
Start process: Lift
Menu item
Only devices with motor lift: In this menu, define the behavior of the flask lift at the
start of the process.
➜
Touch the function button to switch between the two options:
Start process: Lift on
–
moves to the lower end position when starting the process and the evaporation
flask is thus automatically immersed in the heating bath.
Start process: Lift off
–
the upper end position when starting the process; the evaporation flask must
be manually immersed in the heating bath.
➜
Confirm your entry with
Deactivate this function when using a foam brake and on devices
using a large evaporation flask to prevent damage to the flask!
: Touch the function button to activate the function.
: Touch the function button to deactivate the function.
ok
to accept the the change.
: With this selection, the flask lift automatically
: With this selection, the flask lift remains in
ok
.
116