Return-Ground-Guard Connections - Phenix Technologies 400P User Manual

Dc dielectric test set
Hide thumbs Also See for 400P:
Table of Contents

Advertisement

SECTION 5: RETURN-GROUND-GUARD CONNECTIONS
The unit contains a currentmeter feature useful in measurement of different current sources.
1. Return Mode (RTN) (Grounded Return Mode)
This is the standard measurement configuration. The Ground jumper is installed between the Ground
(GND) post and the Return (RTN) post. The low potential side of the test specimen is connected to
Return. It can initially be isolated from ground or tied to ground. When it is connected to Return in this
mode, it will be grounded through the Ground jumper if it was initially isolated from ground.
If the low potential side of the test specimen will not or cannot be isolated from ground, this hookup mode
must be used. Any connections made to Guard must be isolated from ground in this mode and will bypass
the currentmeter. All currents to Ground and Return will be measured in this mode.
2. Guard Mode
In this mode, the Ground jumper is connected to the Guard post and Ground post. The low potential side
of the test specimen must be isolated from ground and connected to the Return post. Only current to the
Return post will be measured by the currentmeter. Any stray leakage currents to ground or from sources
connected to Guard will bypass the currentmeter and not be measured such as stray leakage currents to
ground, stray currents to shields or housings tied to Ground or Guard, etc. In this mode, the low potential
side of the specimen must be isolated from ground. If the low side of the specimen can not or will not be
isolated from ground, this mode cannot be used to accurately measure current flow.
CAUTION: Do not operate unit with the ground jumper removed or not connected to either Return or
Guard. Make sure the Ground post is connected to a good earth ground. See connection diagrams on
next pages.
DCD 4120-10
5-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

4120-10

Table of Contents