INSTALLATION INSTRUCTIONS POWER ON/OFF PROCEDURE POWER SECTION DESCRIPTION TESTING PROCEDURE TROUBLESHOOTING SERVICE AND MAINTENANCE CALIBRATION OF CIRCUIT BREAKER TEST SETS WinHC PC SOFTWARE RETURNED MATERIAL PARTS ORDERING INFORMATION RECOMMENDED SPARE PARTS PARTS LIST ELECTRICAL SCHEMATICS CUSTOMER COMMENTS / SUGGESTIONS HC-20C...
Page 3
WARNING !! Complete Grounding of this unit is necessary for the safe operation of this equipment. Disconnect inputs before un- grounding this equipment HC-20C...
Page 4
DO NOT operate damaged equipment. Remove power, and do not use the equipment until safe operation can be verified by service-trained personnel. Phenix Technologies, Inc. assumes no liability for unsafe or improper use of test equipment. HC-20C...
DUTY CYCLE Continuous 2 Minutes OFF 10 Minutes OFF 30 Minutes OFF HC-12C 12,000 A 2,500 A 1,700 A 1,200 A HC-20C 20,000 A 4,500 A 3,000 A 2,000 A HC-40C 40,000 A 9,000 A 6,000 A 4,000 A HC-75C...
Page 6
Ranges: 0 - 999.999 Seconds 0 - 99,999.990 Cycles Test durations of less than 10 cycles Timer Accuracy: ± 0.004 Seconds ± 0.250 Cycles Test durations of greater then 10 cycles Timer Accuracy: ± 0.010 Seconds ± 0.600 Cycles HC-20C...
SECTION 2: INTRODUCTION Phenix Technologies' High Current Test Sets represent a new era of electrical testing technology. This design is based on time-proven concepts and the power offered by an embedded 32-bit microcomputer. High reliability SCRs have been combined with a phase angle firing system to deliver precisely repeatable current pulses with minimal distortion.
SECTION 3: SAFETY Several safeguards have been built into the Phenix Technologies high current test sets; however, these safeguards are only effective if safe work practices are also used. The test set and these procedures have been designed for use by knowledgeable electrical workers who have been properly trained in its use.
Position the test set in the desired position. For 2-cabinet units, place the regulator cabinet on top of the transformer cabinet. The following procedure is only required for 2-cabinet units (HC-12C and HC-20C): Connect the interconnect ground cable from the regulator cabinet to the transformer cabinet.
Page 10
INSTALLATION INSTRUCTIONS HC-12C, HC-20C, and HC40C INPUT POWER JUMPER CONFIGURATIONS: REGULATOR (TOP) CABINET 220-240V CONNECTION TRANSFORMER (BTM) CABINET 1 2 3 4 220-240V 440-480V 220-240V CONNECTION CONNECTION CONNECTION 380-480V 220-240V CONNECTION CONNECTION DANGER 380-416V CONNECTION INPUT VOLTAGE SELECT REMOVE POWER FROM UNIT...
Switch CB202 (vernier circuit breaker) OFF. Switch the control power OFF (SW201). De-energize the facility power distribution breaker. Disconnect the test set as required. WARNING: Connecting or disconnecting the controller with power applied may damage test set and/or controller. HC-20C...
An SCR Trigger board, PCB201, turns the high current output ON when signaled by the HC Controller. The trigger board receives its command from the HC Controller via an I/O interface PC board, PCB202. The HC Controller precisely controls the phase angle and duration of the output waveform. HC-20C...
Page 13
Normally Open test modes by the test set. If the continuity connections cannot be used (i.e. a single-phase breaker), the set must be operated in the Current Latch mode. CAUTION: DO NOT CONNECT CONTINUITY TERMINALS TO ENERGIZED SOURCE. DAMAGE TO CONTROLLER WILL OCCUR! HC-20C...
Page 14
Connect between the Chassis Ground Terminal and the facility Ground Cable ground. Output Stabs Connection to the circuit breaker under test. Continuity Binding Post For use in contact mode, i.e. Normally Closed or Normally Open. Regulator Speed Control Controls the vernier speed. HC-20C...
Twisting the cables will also help to reduce noise on the current measurement. SAVING RESULTS After a test is performed and while test results are still displayed, click the Save Result button. HC-20C...
Page 16
This mode is used to maintain a preset current value for a long period of time. Click Settings in Test Parameters to access the test parameters. See the specific procedures for the different Test Modes on the next few pages. HC-20C...
Page 17
Click Settings in Test Parameters. Enter the number of cycles that the output will be energized. This should be long enough to get a good reading, but short enough to avoid overheating. Phenix Technologies recommends a value of 7 cycles.
Page 18
This should be long enough to get a good reading and at least twice the expected instantaneous trip time, but short enough to avoid overheating. Phenix Technologies recommends a value of 7 cycles. Be sure to click OK to record the setting.
Page 19
Click the START button to energize the output. Click the STOP button to stop the test at any time. The test set will hold the current until the breaker trips or the operator interrupts the test. Optional – click Save Results. HC-20C...
Page 20
The test set will automatically increase the current up to the HOLD AMPS setting and remain ON until either the operator clicks STOP, the circuit breaker trips, or the MAX DURATION or MAX CYCLES is reached. Click the STOP button to stop the test at any time. HC-20C...
Check to see that main circuit breaker is on. Verify that the breaker is reset and not in the trip position. Verify continuity of the breaker contacts. Check LED indicator on SCR trigger board. It should light when initiated. Check voltage out of SCR's when initiated. HC-20C...
Remove the front cover on the main circuit breaker and clean the main contacts. Replace the contacts if they are beyond repair. Inspect the variable transformer brushes for wear. Replace if they appear worn. Perform basic operational tests to determine proper operation. HC-20C...
This is a self calibrating function, when selected the variac will automatically run from limit to limit to adjust the systems 0% & 100% position sensors. Calibrate Current Meter: When selected you will get the following screen. Enter password 8118. After password is entered you will see the following screen. HC-20C...
Page 24
NOTE: For the 100 kA range your standard will most likely be a much smaller value than your range. In this case you should calibrate near 100% of your standard’s value. The standard should be at least 20% of range. HC-20C...
Page 25
For this calibration you will inject voltages into the banana jacks over the output panel. (NOT CONTINUITY BINDING POSTS) Raise the voltage source to approximately 80% of range. Enter the value from your standard in the Actual Voltage field and press update. The Cal Factor will automatically update. Repeat for other ranges. HC-20C...
Page 26
It is recommended that documentation of the cal factors be maintained, in the event that the cal factors become corrupt or if inadvertently changed and need to be restored. This can be done by going to the Hardware Settings screen. HC-20C...
SECTION 11: WinHC PC Software About WinHC Welcome to WinHC, circuit breaker test system software developed by Phenix Technologies to control the Phenix Technologies high current test sets. WinHC provides advanced database and report generation capabilities, enabling you to save your circuit breaker test results in a database for future recall and report generation.
Page 28
Vernier Control: The position of the vernier determines the level of the output voltage and current of the WinHC. Click the up/down arrows to change the position of the vernier, or enter a vernier position and click Go... HC-20C...
Page 29
You MUST NOT enter a desired test current that is at or above the level where the breaker under test will trip. If the breaker trips during initialization, then the process will be aborted. HC-20C...
Page 30
Use the Output Control screen to perform the circuit breaker test. When the test is complete, click Save Result to save the test results. Click Waveform... to view the waveform of the output current. HC-20C...
Page 31
NOTE: The search will be performed on the field selected by Sort By…. 4. Click Test Report… to generate a report. You will be given the option of pre-viewing the report on screen or sending it directly to the printer. HC-20C...
Page 32
WinHC data to a network server. You are responsible for moving the data files, if you change this setting without moving the files, WinHC will create empty data files in the new location. HC-20C...
Page 33
The labels options screen contains the field labels for the optional circuit breaker nameplate fields. You may change any of these to suit your particular needs. These labels appear on the circuit breaker nameplate screen, under Additional Info…, and on test reports. Lists The Lists Options HC-20C...
Page 34
Tests The Tests Options Circuit breaker Test Titles: You may customize the test titles to match your individual testing procedures. Test Titles appear on the test screen and print on test reports. HC-20C...
Page 35
The backup data screen allows you to easily select a drive/directory as a destination for your data backup. Access this screen by selecting Tools and then Backup Data…. Select the drive/directory and Click Backup Data to begin the backup. HC-20C...
Page 36
The restore data screen allows you to easily select a drive/directory as a source for your data recovery. Access this form by selecting Tools and then Restore Data…. Select the drive/directory and Click Restore Data to begin the data recovery. HC-20C...
Serial Number Reason for Return Cause of Defect If Phenix Technologies, Inc. deems return of the part appropriate, it will then issue an "Authorization for Return." If return is not deemed advisable, other inspection arrangements will be made. NOTE: Material received at this plant without the proper authorization shall be held as "Customer's Property"...
When your purchase order is received at our office, a representative of Phenix Technologies will contact you to confirm the current price of the part being ordered. If a part you order has been replaced with a new or improved part, an Applications Engineer will contact you concerning any change in part number.
14-1 SECTION 14: RECOMMENDED SPARE PARTS HC-20C In order to maintain your set in full operating condition with a minimum of down time, the following spare parts should be kept on hand to avoid unnecessary phone calls, expensive modes of shipment, delays in repairs, etc.
17-1 SECTION 17: CUSTOMER COMMENTS / SUGGESTIONS Phenix Technologies made significant efforts to ensure that the materials in this Operator’s Manual are correct. If there are concerns or comments as you have used this information, Phenix Technologies appreciates any feedback.
Need help?
Do you have a question about the HC-20C and is the answer not in the manual?
Questions and answers