Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Diffustik resp. the variant PFTstik could be damaged.
Reason: Electrostatic discharges. Therefore:
Attention
Version: 8 | Release date: 16. August 2021
Possible danger to life.
Reason: Electric shock due to lack of galvanic
separation with composition of non-approved
components. Therefore:
•
Only use the USB connection cable
supplied by the manufacturer as spare parts,
which are always part of the medical device!
•
The device may only be operated in conjunction
with the equipment cart (ergoline ergocar PC)
offered by manufacturer!
Reason: Electrical shock and / or misdiagnosis due to
loss of electrical safety caused by exceeding the
recommended maintenance schedule. Therefore:
•
Regularly check the specified maintenance
schedule!
•
If a maintenance schedule is exceeded, do not
continue to use the total system! Request
maintenance work from your authorised
specialist retail partner!
Reason: Misdiagnosis caused by measurement error.
Therefore:
•
Do not connect any additional USB devices!
(except mouse, keyboard and printer)
•
Do not install any other software!
•
Ask your authorised specialist retail partner
which devices are approved by the
manufacturer!
•
Preferably no floor made of synthetic material!
•
Otherwise a relative air humidity of at least
30 % is required!