User's Requirements
EN60825 (IEC60825)
Requirements;
Sub-clause
Class 1
Remote interlock
Not required
Key control
Not required
Beam attenuator
Not required
Emission
Not required
indicator device
Warning signs
Not required
Beam path
Not required
Specular
No requirements
reflection
Eye protection
No requirements
Protective
No requirements
clothing
Training
No requirements
With respect to the requirements of remote interlock connector, key control, beam attenu-
ator, and emission indicator, Class 3B laser products not exceeding five times the AEL of
Class 2 in the wavelength range of 400 to 700 nm are to be treated as Class 3A laser
products.
Note: This table is intended to provide a convenient summary of requirements. See text of
JIS C6802
Item
Class 1
Remote
Not required
interlock
Key control
Not required
Beam breaker
Not required
or attenuator
Warning sign
Not required
Beam path
Not required
Mirror
Not required
reflection
Eye protection Not required
Protection
Not required
clothes
Training
Not required
Note: *Class 3B applies to any laser beam with a power of 5 mW maximum in the visible
Classification
Class 2
Terminate beam at end of useful length
Required if engineering and administrative procedures
not practicable and MPE exceeded
Required for all operator and maintenance personnel
standard for complete precautions.
Class 2
Class 3A
The laser beam must be terminated and, as a rule, must be enclosed. If the laser
beam is exposed, the vertical height of the beam must not be the same as that of
the eyes.
Wear protection clothes if exposure of the skin to the laser beam
may exceed the MPE of the skin.
The laser system must be operated by only properly trained peo-
ple.
range of the laser beam.
Class 3A
Class 3B
Connect to room or door circuits
Remove key when not in use
When in use prevents inadvertent
exposure
Indicates laser in energized
Follow precautions on warning signs
Prevent unintentional reflections
Sometimes
required
Class 3B
3B*
3B
Connect the remote interlock of
the laser beam to the emer-
gency main interlock, the inter-
lock of the room, or the
interlock of the door.
Do not keep the key in the lock
when the laser beam is not
used.
Used to protect people from
accidental radiation by the laser
beam.
Post a proper warning sign on the door to the
room where laser beam equipment is installed.
Appropriate optical elements must be securely
attached and you must be able to control the
optical elements during laser radiation.
Use eye protectors except in
special, specified locations.
SECTION 4
Appendix
Class 4
Specific
requirements
Class 4
Z500