4.2.14 Complete

When [VTBI] is not set during the infusion and infusion is completed, if drop
sensor is installed and the switch of [Drop sensor] is on, the [Empty] alarm will
be triggered.
When [VTBI] is set during the infusion and the remaining infusion time is close to
the [Time Near End] set by the users, the [Time Near End] alarm will be triggered.
If no action has been taken, the alarm will not be cancelled automatically until the
infusion is completed, and then switch to [VTBI Complete] or [Empty] alarm. Set
[Time Near End], please refer to 6.9 Time Near End. When infusion is completed,
enter to [KVO] mode, and KVO mode will run for 30 mins at most. Infusion will
stop automatically after the KVO is finished, and the [KVO Finish] alarm will be
triggered. Set KVO rate, please refer to 6.1 KVO.

4.2.15 Standby

Under non-running status, press
interface. Press
pump cannot be put in standby mode if there is a high-level alarm.
When the standby state is ended, the title bar will display [Standby Time
Expired], press

4.2.16 Turn off the Pump

Follow the steps below to turn off the infusion pump:
1.
Disconnect from the animal;
2.
Hold down
will turn off.
NOTE
When powering off normally, the current operating data and saved data
will be autosaved.
to enter the standby countdown shutdown
to modify standby time (range is 00:01-99:59 h:min). The
or
to cancel alarm.
, until the Turn Off progress bar complete, and the power
4-8