A.4 Specifications
Parameters

Specifications

Infusion set used in conjunction with
infusion pump should meet the
Infusion set
requirements of ISO 8536-4:2004
standard
Infusion equipment for medical use—
Part 4: Infusion sets for single use,
gravity feed, MOD
Unit of Rate (ml/h): 0.1-2000ml/h, the
minimum increment is 0.1ml/h
Unit of Rate (gtt/min): 1-(400*Drip /60)
gtt/min, the minimum increment is
1gtt/min
Note 1: The maximum rate of the
infusion system (infusion pump and
infusion set) is affected by inner and
Rate
outer diameter, material, elasticity of
the infusion set and other factors.
Therefore, infusion sets of different
brand and model may differ in
maximum rates.
Note 2: The above declared rates are
based on BOON A2 and B.Braun
Intrafix Safeset infusion sets.
Bolus Rate
0.1-2000ml/h
Purge Rate
800ml/h, nonadjustable
VTBI
0.1-9999ml
Time
00:00:01-99:59:59 h:m:s
Volume
0.1-9999ml
History Record
Can store up to 1500 history records
Select Mode
Rate Mode, Body Weight Mode, Time
Factory Default
/
---
The previous Bolus Rate is
the default value, if not, the
Bolus Rate will be set as the
below rules:
1. If the maximum rate
≥800ml/h, the default bolus
rate is 800ml/h;
2. If the maximum rate
<800ml/h, the default bolus
rate is the maximum rate.
800ml/h
---
---
0ml
/
Rate Mode
A-5