Parameter Memory Function; Select Infusion Mode; Purge Air - Mindray BeneFusion VP3 Vet Operator's Manual

Infusion pump
Table of Contents

Advertisement

4.2.6 Parameter Memory Function

In clinical treatments, the medical staffs need to initiate the infusion as soon as
possible during emergency situations, infuse the liquid drug into the animal's body
in the shortest time possible, and set the detailed parameters later during infusion.
1. Select [Main Menu]→[User Maintenance]→Input User Maintenance
Password→[Para. Memory].
2. Select [Para. Memory]→[On]. If [Off] is selected, the following steps cannot
be conducted.
3. After selecting the infusion set brand, the previous infusion screen will appear,
the previous therapy parameters will be loaded, the users can use the
previous treatment parameters.

4.2.7 Select Infusion Mode

Press
to enter [Main Menu]→[Select Mode]. On this interface, users can
press
Infusion Mode for detailed introduction of each infusion mode.

4.2.8 Purge Air

During infusion, the user should prevent air bubbles from entering the blood with
the liquid drug, which may form an aeroembolism and put animals in serious
danger. Therefore, air bubbles in the infusion tube should be eliminated before
the infusion. Under non-running status of any infusion mode, press
[Purge Air] prompt screen. Hold down
release
after the air bubbles are purged.
WARNING
Please disconnect the pump from animal before purge. Otherwise, the
animal will be in serious danger!
and
to select infusion mode. Please refer to Chapter 5
to enter [Purge Air] running screen,
4-5
to enter

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BeneFusion VP3 Vet and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF