Mindray BeneFusion VP3 Vet Operator's Manual page 78

Infusion pump
Table of Contents

Advertisement

RATED power
30 A/m
frequency
50 Hz / 60 Hz
magnetic fields
IEC 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the specified electromagnetic environment. The
customer or the user of the device should assure that it is used in such an environment
as described below.
IEC 60601 Test
Immunity test
level
Conducted
3 Vrms
disturbances
150 kHz to 80
induced by RF
MHz
fields
6 Vrms
IEC61000-4-6
in ISM bands
and amateur
radio bands
between 0,15
MHz and 80
MHz
Radiated RF
10V/m
EM fields
80 MHz to 2.7
IEC61000-4-3
GHz
27 V/m
Proximity fields
380–390 MHz
from RF
wireless
28 V/m
communication
430–470 MHz,
s equipment
800–960 MHz,
IEC61000-4-3
1700–1990
MHz,
2400–2570
MHz
30 A/m
50 Hz / 60 Hz
Compliance
level
10 Vrms
6 Vrms
a
10V/m
27 V/m
28 V/m
B-4
power supply or a
battery.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Electromagnetic
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter. Recommended
separation distance:
3
5 .
150k to 80 MHz
d
P
 
 
V
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
environment
-