Supported Device ............................1 - 4 ™ Spectrum OR ............................1 - 4 Viewstation OR Main Screen ..................... 2 - 1 The Main Screen ............................2 - 1 Waveform Parameter Tiles........................2 - 2 Waveform Area ..........................2 - 3 Waveform Digital Area ........................
Viewstation OR. For additional information and technical assistance, please contact a Mindray DS Patient Monitoring Representative. A general understanding of the features and functions of the Viewstation OR and its components (shown in FIGURE 1-1) is a prerequisite for the proper use of this equipment.
Page 9
WARNING: Route cables neatly. Ensure cables are not in the way of patient or hospital personnel. WARNING: All of the equipment on the Viewstation OR network must utilize the hospital emergency power system. Failure to do so will result in loss of monitoring during extended periods of power failure.
Page 10
The Viewstation OR must be installed outside the patient’s vicinity. CAUTION: The Viewstation OR CPU should not be used within 25 cm of the point where leakage or discharge of a flammable anesthetic mixture with oxygen or nitrous oxide may occur.
Introduction Indications for Use Indications for Use The indications for use for the Viewstation OR include viewing real time patient clinical data. ™ The clinical data displayed by the Viewstation OR is obtained from a Spectrum OR patient monitor and includes: ECG waveforms, Invasive and Non-Invasive Blood Pressures, Blood...
Touch Screen Direct Current (DC) Earth Ground Patient Panorama Patient Network Keyboard Network Ethernet Mark of compliance with U.S. Mouse and Canadian safety standards requirement. Do not dispose of this product as Input/Output unsorted municipal waste. Viewstation OR™ Operating Instructions 0070-10-0663-01...
Page 13
Introduction Symbols and Descriptions This page intentionally left blank. 0070-10-0663-01 Viewstation OR™ Operating Instructions...
System Overview This section provides a high level view of the authorized components for the Viewstation OR. • The Viewstation OR • Basic Configuration • Supported Device Viewstation OR™ Operating Instructions 0070-10-0663-01 1 - 1...
(See section 1.2, Basic Configuration.) While accessing the tabs that are associated with the menu bar buttons, the Viewstation OR display automatically re-configures as a split-screen. The top half of the screen displays patient data while the bottom half of the screen is converted to a work space for simultaneously entering and modifying display configurations.
Page 16
8 PORT TO PANORAMA ETHERNET SWITCH OR GATEWAY VIDEO SPLITTER EXTERNAL EXTERNAL EXTERNAL TOUCH SCREEN DISPLAY DISPLAY DISPLAY TOUCH PAD TOUCH PAD FIGURE 1-2 Example Multiple Display Configuration of the Viewstation OR Viewstation OR™ Operating Instructions 0070-10-0663-01 1 - 3...
1.3.1 Spectrum OR All Viewstation OR data is sourced from the Spectrum OR, a vital signs bedside monitor designed specifically for the operating room environment. The following table lists the Spectrum OR parameters that are supported by the Viewstation OR.
Viewstation OR Main Screen This section provides an overview of the Viewstation OR main screen. The Main Screen The Main Screen for the Viewstation OR is used to display patient data and to navigate the system. The Main Screen includes the following elements: •...
The Main Screen Viewstation OR Main Screen 2.1.1 Waveform Parameter Tiles Waveform parameter tiles are the core elements of the Viewstation OR system (see ™ FIGURE 2-1). These tiles display patient data obtained from the Spectrum OR . Three to eight waveform parameter tiles may be displayed at once.
IBP area has been labeled at the Spectrum OR, the label will also display in the Viewstation OR. Wave gain settings include 0 – 20, 0 – 40, 0 – 80, 60 – 140, 0 – 160, 0 – 225, 0 – 320, and (-10) – 10 mmHg.
The IBP1 through IBP4 waveform digital areas display Systolic, Diastolic and Mean blood pressure numeric values. When an IBP area has been labeled at the Spectrum OR, it will also display in the Viewstation OR. Labels include ART, ICP, CVP, LA, LV, RA, PA and UA.
Viewstation OR Main Screen The Main Screen 2.1.2 Digital Parameter Tiles The digital parameter tiles display up to six parameters, based on the sensors available for ™ the attached Spectrum OR . Below is a list of available digital parameter tiles.
Page 23
The Main Screen Viewstation OR Main Screen FIGURE 2-9 Example PVC/min Digital Configuration Dialog NOTE: The Change To choice control is based on the sensors utilized ™ by the Spectrum OR 2. Select the Done or Cancel button. • The Done button accepts the changes.
The Main Screen 2.1.3 Menu Bar The menu bar provides navigation for the Viewstation OR and is fixed at the bottom of the screen. The table below provides a list of the functions and tabs associated with each menu button.
Messages are shown in all bold text. 2.1.5 Date and Time The date and time, which are provided in the bottom right corner of the Viewstation OR display, are formatted during software installation. FIGURE 2-11 Date and Time 2 - 8 0070-10-0663-01 Viewstation OR™...
Viewstation OR Main Screen The Main Screen 2.1.6 Display States The Viewstation OR displays full-screen messages when it has left active display mode. The messages and possible resolutions are described below. MESSAGE REASON RESOLUTION Standby The Spectrum OR is in a standby Remove the Spectrum OR from the state.
Page 27
The Main Screen Viewstation OR Main Screen This page intentionally left blank. 2 - 10 0070-10-0663-01 Viewstation OR™ Operating Instructions...
The active display configuration name is listed in the black text box. An asterisk appears to the left of the display configuration name when unsaved changes have been made. FIGURE 3-2 Example Waveform Parameter Tile Buttons FIGURE 3-3 Example Digital Parameter Tile Buttons 3 - 2 0070-10-0663-01 Viewstation OR™ Operating Instructions...
Page 30
Display Tab Display Configuration Label Dialog Up to 10 pre-configured displays are available on the Viewstation OR. The Display Configuration Label dialog lists all of the saved Viewstation OR display configuration labels. The following factory default settings and their associated parameters may be altered as needed during initial setup: •...
An asterisk appears next to the configuration name, indicating that unsaved changes have been made to that configuration. 3. Repeat steps 1 and 2 for all tiles to which a new parameter will be assigned. 3 - 4 0070-10-0663-01 Viewstation OR™ Operating Instructions...
Page 32
Are you sure you want to exit this tab without saving?” Select Yes to exit the Display tab without saving any changes. Select No to close the dialog box and return to the Display tab. FIGURE 3-5 Exiting without saving Viewstation OR™ Operating Instructions 0070-10-0663-01 3 - 5...
FIGURE 3-9). Enter the new configuration name and select the Enter button. (Or select the Close button to return to the Display tab without saving the new configuration.) • Select the Cancel button to return to the Display tab without making changes. 3 - 6 0070-10-0663-01 Viewstation OR™ Operating Instructions...
Display tab without editing the display.) 3. Edit the label name and select the Enter button or select the Cancel button to return to the Display tab without editing the display name. Viewstation OR™ Operating Instructions 0070-10-0663-01 3 - 7...
Page 35
A Backspace (<==) button to clear a single character in the edit box • A Caps Lock button to enable or disable caps lock (A Caps Lock indicator is shown when Caps Lock is enabled.) 3 - 8 0070-10-0663-01 Viewstation OR™ Operating Instructions...
A display configuration cannot The configuration is the current Close the display and then be removed. display. remove the configuration. The configuration is not The configuration has not been Save the configuration. displayed. saved. Viewstation OR™ Operating Instructions 0070-10-0663-01 3 - 9...
FIGURE 3-10 Example Wave Gain Tab 3.2.2 Configuring Wave Gain Settings Viewstation OR waveform tiles will initially display the default settings as described in "Wave Gain Tab (System)" on page 4-5. Configurations made in Patient Setup mode only affect the active session.
The wave gain settings did not The Done button was not Adjust the wave gain settings change selected after the wave gain and select the Done button settings were changed. when the settings are completed. Viewstation OR™ Operating Instructions 0070-10-0663-01 3 - 11...
Page 39
Wave Gain Tab (Patient) Patient Setup Functions This page intentionally left blank. 3 - 12 0070-10-0663-01 Viewstation OR™ Operating Instructions...
This section outlines assigning a color to a parameter in the Parameter Color tab (see FIGURE 4-2). 1. Select a color tile button. A checkmark indicates that the colored tile is selected. FIGURE 4-2 Example of a selected color tile 4 - 2 0070-10-0663-01 Viewstation OR™ Operating Instructions...
The Equipment Setup tab connects and disconnects the Spectrum OR from the Viewstation OR computer. The Device ID text box lists the device ID of the Spectrum OR that is currently connected to the Viewstation OR. FIGURE 4-3 Example Equipment Setup Tab Viewstation OR™...
SOLUTION The data that is being displayed at the A different Spectrum OR Connect the desired Viewstation OR does not correspond monitor than the one that is Spectrum OR monitor to to the data that is being displayed at desired is being displayed the Viewstation OR.
40, 60, and 100 mmHg. 4.3.3 Sidebar Button - Normal Screen Select the Normal Screen sidebar button to return the display to the full screen mode, and close all tabs, message boxes, and dialog boxes. Viewstation OR™ Operating Instructions 0070-10-0663-01 4 - 5...
System Setup Functions Touch Screen Tab A Viewstation OR configuration can include a touch screen display. A touch pad or mouse can also be used for point-and-click system navigation. This section provides an overview of how to turn the touch screen on and off and how to recalibrate it.
3. The message Touch the screen. Does the cursor follow your finger? is displayed (see FIGURE 4-9). Touch all four corners and all four sides of the screen. FIGURE 4-9 Check Calibration Screen Viewstation OR™ Operating Instructions 0070-10-0663-01 4 - 7...
Contact your local Mindray DS Service Representative if recalibration of the touch screen does not improve the performance of the touch screen display. 4 - 8 0070-10-0663-01 Viewstation OR™ Operating Instructions...
User Maintenance This section of the manual outlines proper maintenance of the Viewstation OR touch screen. Viewstation OR™ Operating Instructions 0070-10-0663-01 5 - 1...
Care and Cleaning of the Touch Screen User Maintenance Care and Cleaning of the Touch Screen For the best performance, it is recommended that the Viewstation OR touch screen be kept clean. NOTE: The touch screen must be turned off prior to cleaning. (See "Turning the Touch Screen On and Off"...
Parameter Specifications This section is for reference only to list the allowable range specifications for all of the ™ parameters measured by the Spectrum OR Viewstation OR™ Operating Instructions 0070-10-0663-01 6 - 1...
Neonate Pediatric Adult The Viewstation OR displays ECG waveforms when using 3-wire and 5-wire lead cables. The ECG leads available at the Viewstation OR are dependent on the lead cable connected to the patient. LEAD WAVEFORM LEADS AVAILABLE 3-wire lead set...
Respiration Source: CO PATIENT SIZE MINIMUM VALUE (rpm) MAXIMUM VALUE (rpm) Neonate Pediatric Adult The Viewstation OR supports respiration waveforms. NIBP The NIBP parameter includes measurements for the Systolic (Sys), Diastolic (Dia), Mean and Elapsed Time (ET). Unit of Measure: mmHg Resolution:...
The Viewstation OR supports SpO waveforms. The Viewstation OR supports up to four invasive blood pressure parameters. The measurements shown for each pressure include Systolic (Sys), Diastolic (Dia) and Mean. IBP parameter labels include IBP, Art, UA, LV, PA, CVP, NOTE: ICP, LA, and RA.
Pediatric Adult The Viewstation OR supports IBP waveforms. The Viewstation OR will not display an IBP waveform if the IBP cable is not connected to the system. The Viewstation OR supports the mmHg, kPa and % units of measure for CO...
Parameter Specifications Invalid Data Display 6.11 Invalid Data Display The Viewstation OR uses dashes (---), Xs (XX) or a straight line (waveform) to show invalid data. Dashes (---) Digital Data Digital data is replaced with dashes (---) for the following reasons: •...
Page 57
Invalid Data Display Parameter Specifications This page intentionally left blank. 6 - 8 0070-10-0663-01 Viewstation OR™ Operating Instructions...
Performance Specifications Viewstation OR Computer Viewstation OR Computer There is one basic hardware platform for the Viewstation OR computer system. It contains the following: • 1 GB of RAM • 80 GB system hard drive, minimum • Four channel serial interface for touch screen inputs and future features •...
• 1060 to 700 hPa (-1250 to 9889 feet ASL) (-380 - 3014 m) (795 to 525 mm Hg) 7.7.9 Safety Standards The Viewstation OR complies with relevant portions of the following standards: 1. UL 60950-1:2003 Safety of Information Technology Equipment 2.
GUIDANCE AND MINDRAY DS USA INC. DECLARATION - ELECTROMAGNETIC EMISSIONS The Viewstation OR is intended for use in the electromagnetic environment specified below. The customer or the user of the Viewstation OR should assure that it is used in such an environment. ELECTROMAGNETIC EMISSIONS TEST...
Page 65
GUIDANCE AND MINDRAY DS USA INC. DECLARATION - ELECTROMAGNETIC IMMUNITY The Viewstation OR is intended for use in the electromagnetic environment specified below. The customer or the user of the Viewstation OR should assure that it is used in such an environment. IMMUNITY IEC 60601...
Page 66
GUIDANCE AND MINDRAY DS USA INC. DECLARATION - ELECTROMAGNETIC IMMUNITY The Viewstation OR is intended for use in the electromagnetic environment specified below. The customer or the user of the Viewstation OR should assure that it is used in such an environment. IMMUNITY IEC 60601...
COMMUNICATIONS EQUIPMENT AND THE VIEWSTATION OR AND TELEMETRY SERVER The Viewstation OR is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of thee Viewstation OR can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Viewstation OR as recommended below, according to the maximum output power of the communications equipment.
I/O Control Devices I/O Control Devices The Viewstation OR can be used with a mouse, touch pad and touch screen. These I/O control devices are used to make selections from various menus and screens. The mouse and touch pad are supported for system diagnostics and user interface, operating independent of the touch screen.
Warranty Domestic Warranty Mindray DS USA Inc. warrants that its products will be free from defects in workmanship and materials for a period of one (1) year from the date of purchase. Disposable or one-time use products are warranted to be free from defects in workmanship and materials up to a date one year from the date of purchase or the date of first use, whichever is sooner.
Page 70
Performance Specifications Warranty Mindray DS USA Inc. shall not be liable for any incidental, special, or consequential loss, damage, or expense directly or indirectly arising from the use of its products, liability under this warranty and the buyer’s exclusive remedy under this warranty is limited to servicing or replacing at Mindray DS’s option at the factory or at an authorized Mindray DS Distributor,...
Please include the instrument model number, the serial number, and a description of the problem with all requests for service. Any questions regarding the warranty should be directed to the nearest Mindray DS location. A list of international offices, along with their phone numbers, is provided at the end of this manual.
Page 72
Glossary This section provides a glossary of terminology used throughout the Viewstation OR Operating Instructions. • Terms, Acronyms, and Abbreviations Viewstation OR™ Operating Instructions 0070-10-0663-01 8 - 1...
Page 73
Terms, Acronyms, and Abbreviations TERMS DEFINITIONS Bedside Monitor A monitoring device that has its own user interface and does not require the Viewstation OR to be its primary monitoring source Continuous Cardiac Index Continuous Cardiac Output Cardiac Index Cardiac Output...
Page 74
Right Arm when referring to ECG Leads (AHA) RESP Respiration Rate Right Leg when referring to ECG Leads (AHA) Respirations per minute Surface Acoustical Wave Temperature of the Blood Umbilical Artery Chest Lead (AHA) Viewstation OR™ Operating Instructions 0070-10-0663-01 8 - 3...
Page 75
0070-10-0663-01 Rev C September 06, 2010...