ICS ELECTRONICS 4863 Manual page 50

Table of Contents

Advertisement

e.g.
SYST:COMM:GPIB:ADDR:EXT 1
*SAV 0
The 4863 reads the external address switch at power turn-on time. Always
cycle power off and back on after changing the address switch setting. On
ICS's GPIB Connector/Address Switch Assemblies, the rocker switches are
laid out as shown below in Figure 2-8.
Notes: Switch show set to address 4
Figure 2-8
2.7.3.5 Board Jumpers
4863/2363s have two jumpers on the PC board as shown in Figure 2-6. The
jumper functions are listed in Table 2-6
TABLE 2-6
Jumper
W1
Default - Returns the unit to its factory default
settings when in place at power turn-on time.
Leave out for normal operation
W2
Write Enable - Must be in place to write to
or save in flash memory.
1 8 4 2 1 S
- -
6
Rocker 'S' used for Address 0
Address Switch Layout
4863/2363 JUMPER SETTINGS
Functions
2-21
1
0
Factory
Setting
Omitted
Installed
2