5. Specifications
The pipettor is a high quality instrument which offers
excellent accuracy and precision. The accuracy and precision
(repeatability) of the liquid volume depends on the quality
of pipet tips used. The values for accuracy and precision
shown in the table below were obtained using manufacturer
non-filter pipet tips. Those tips are recommended for use to
ensure compatibility, accuracy and precision when pipetting.
Nominal
Volume (µL)
2
5
10
20
40
50
70
75
100
120
150
200
250
300
400
500
1,000
The accuracy and precision were obtained gravimetrically
using manufacturer tips performing at least 10 measurements
of distilled water at a temperature of 20°C ± 1°C according to
EN ISO 8655 standards.
The use of tips from other manufacturers or filter tips may
result in incorrect liquid aspiration and require pipettor
recalibration.
The design of the pipettor enables the user to recalibrate it
according to the information presented in Section 9.
4 – LABNET LABPETTE FX PIPETTORS
Cat. No.
P3900-2
P3900-5
P3900-10
P3900-20
P3900-40
P3900-50
P3900-70
P3900-75
P3900-100
P3900-120
P3900-150
P3900-200
P3900-250
P3900-300
P3900-400
P3900-500
P3900-1000
Accuracy
(%)
±3.0
±2.0
±1.2
±0.9
±0.9
±0.9
±0.9
±0.9
±0.9
±0.8
±0.7
±0.6
±0.6
±0.6
±0.6
±0.6
±0.6
Precision
(%)
≤1.5
≤1.0
≤0.5
≤0.4
≤0.3
≤0.3
≤0.3
≤0.3
≤0.3
≤0.3
≤0.3
≤0.2
≤0.2
≤0.2
≤0.2
≤0.2
≤0.2
Non-filter
Tips (µL)
200
200
200
200
200
200
200
200
200
200
200
200
1,000
1,000
1,000
1,000
1,000
Need help?
Do you have a question about the Labpette FX and is the answer not in the manual?
Questions and answers