Readhead mounting and alignment

Mounting brackets
The bracket must have a flat mounting surface and should provide adjustment to enable conformance to the
installation tolerances, allow adjustment to the rideheight of the readhead, and be sufficiently stiff to prevent
deflection or vibration of the readhead during operation.
Readhead set-up
Ensure that the scale, readhead optical window and mounting face are clean and free from obstructions.
NOTE: When cleaning readhead and scale apply cleaning fluid sparingly, do not soak.
To set nominal rideheight, place the green spacer with the aperture under the optical centre of the readhead
to allow normal LED function during set-up procedure. Adjust the readhead to maximise the signal strength
along the full axis of travel to achieve a Green set‑up LED on the readhead (> 70% signal). If a digital Ti / TD
interface is used, aim for a Blue LED on the interface.
NOTE: The readhead should be installed and set-up with the AGC switched off (CAL LED off).
When re-installing, factory defaults should be restored
0° ±0.4°
Pitch
0° ±1°
TONiC RKLC20-S linear installation guide
('Restoring factory defaults', page
Readhead set-up LED status
Yaw
Green
Orange
Roll
0° ±0.5°
Green spacer
Rideheight
2.1 ±0.15 mm
T103x readhead LED diagnostics
LED
Incremental
Set-up
Reference
Operating
15).
CAL
Calibration
*
 Flash will effectively be invisible when incremental signal is > 70% when passing reference mark.
Red
Ti0004 to Ti20KD and TD4000 to TD0040 interface LED diagnostics
Signal
Incremental
Reference
mark
Alarm output will take the form of 3‑state or line driven E− signal depending on interface configuration.
Also, some configurations do not output overspeed alarm.
Momentary status only, while fault condition remains.
Alarm may result in axis position error, re-datum to continue.
See the TONiC encoder system Data sheet (Renishaw part no. L‑9517‑9337) for interface configuration details.
This can be downloaded from our website at
Renishaw representative.
Indication
Green
Normal set‑up: signal level > 70%
Orange
Acceptable set-up; signal level 50% to 70%
Poor set-up; signal may be too low for reliable
Red
operation; signal level < 50%
Green (flash)*
Normal phasing
Orange (flash)
Acceptable phasing
mark
Red (flash)
Poor phasing; clean scale and recalibrate if required
On
Automatic Gain Control – On
Off
Automatic Gain Control – Off
Single flashing
Calibrating incremental signals
Double flashing
Calibrating reference mark
Flashing at
Reset
Restore factory defaults
power‑up (< 2s)
Indication
Purple
Normal setup; signal level 110% to 135%
Blue
Optimum setup; signal level 90% to 110%
Green
Normal set-up: signal level 70% to 90%
Orange
Acceptable set-up; signal level 50% to 70%
Poor set-up; signal may be too low for reliable
Red
operation; signal level < 50%
Red / blank -
Poor set‑up; signal level < 20%; system in error
flashing
Blue / blank -
Over speed; system in error
flashing
Purple / blank -
Over signal; system in error
flashing
Blank flash
Reference mark detected (speed < 100 mm/s only)
www.renishaw.com/tonicdownloads
Status
Alarm
Status
output
No
No
No
No
No
Yes
Yes
Yes
No
and is also available from your local
13