Restoring Factory Defaults; Switching Automatic Gain Control (Agc) On Or Off - Renishaw TONiC T103x RKLC20-S Installation Manual

Linear encoder system
Table of Contents

Advertisement

Restoring factory defaults

When realigning the readhead, reinstalling the system, or in the case of continued calibration failure, factory defaults should be restored.
To restore factory defaults:
Switch system off.
u
u
Press and hold the CAL button whilst switching the system on. The CAL LED on the readhead will flash several times, indicating that the factory defaults have been restored.
u
Release CAL button.
u
Check the
'Readhead mounting and alignment', page 13
NOTE: System must be re-calibrated after restoring factory defaults.

Switching Automatic Gain Control (AGC) on or off

AGC can be switched on or off via the interface.
Press and hold the CAL button on the interface for > 3 seconds to switch AGC on or off. The CAL LED on the readhead will be illuminated when AGC is active.
u
NOTE: The system must be calibrated before switching AGC on
TONiC RKLC20-S linear installation guide
and recalibrate the system
('System calibration', page
('System calibration', page
14).
14).