Additional Safety Requirement - Csa Only; Certifications And Approvals; Batteries - Teledyne Everywhereyoulook GT-FIRE User Manual

Portable gas detector
Table of Contents

Advertisement

GT-FIRE
PORTABLE GAS DETECTOR
WARNING: TO REDUCE RISK OF EXPLOSION, DO NOT MIX OLD
BATTERIES WITH USED BATTERIES OR MIX BATTERIES FROM DIFFERENT
MANUFACTURERS.
WARNING: NEVER ATTEMPT TO RECHARGE NON-RECHARGEABLE
CELLS.
Caution: Not for use in oxygen enriched atmospheres.
1.2.1. Additional Safety Requirements - CSA Only
Caution: Before each day's usage, sensitivity must be tested on a
known concentration of Methane equivalent to 25% - 50% of full-scale
concentration. Accuracy must be within 0 to +20% of actual. Accuracy
may be corrected by using the GT Calibration Software and manually
applying the gas.
1.2.2. Certification and Approvals
The monitor has the following approvals:
UL 913 Class I, Div 1 Groups C and D.

1.2.3. Batteries

Caution: Only use approved batteries.
1. UL APPROVED MONITORS:
A. Alkaline: Any 'LR14' type
B. Rechargeable NiMH: Any 'C' type
2. ATEX/IECEx/CSA APPROVED MONITORS:
A. Alkaline:
i.
Energizer No. E93, Alkaline, (Zn/MnO2), LR14 Size 'C' cell
ii. Duracell Procell, Alkaline, (Zn/MnO2), LR14 Size 'C' cell
iii. Duracell Plus, Alkaline, (Zn/MnO2), LR14 Size 'C' cell
iv. Duracell, Alkaline, (Zn/MnO2), LR14 Size 'C' cell
B. Rechargeable:
i.
Panasonic Type HHR-2SRE Nickel-Metal Hydride cell
8
Mark
67816
Revision D

Advertisement

Table of Contents
loading

Table of Contents