2. To enter/edit the gas cylinder concentration, press and hold the NEXT
gas value to be altered (as shown in
DOWN
buttons as necessary to enter the correct gas value.
Note: Pressing and holding either of the buttons will rapidly increase/decrease the
gas value.
3. Once the monitor has been programmed with the actual gas cylinder content, press and hold
the OK
button to start the test. The screen shown in
be displayed.
• Left column - gas ranges available in the monitor
• Middle column - concentration of gas being applied
• Right column -peak gas reading
4. Apply the gas using a demand flow regulator.
5. Wait for the "tick" symbol to appear next to applied range (as shown in
- In Progress &
Completed).
Figure 56: Bump Test - In Progress & Completed
6. Either exit the Bump Mode and store the values automatically or go back to the CalGas menu
to define and apply gas from another cylinder.
• Exit the Bump Mode - press and hold the EXIT
• Return to CalGas selection display - press and hold OK
67816
Revision D
Figure 54: Enter/Edit CalGas
Figure 54: Enter/Edit CalGas Value
Figure 55: Applied Gas Peak Values
PORTABLE GAS DETECTOR
button to highlight the
Value). Press the UP
Figure 55: Applied Gas Peak Values
Figure 56: Bump Test
button.
button.
GT-FIRE
or
will
43