GT-FIRE
PORTABLE GAS DETECTOR
3. Select NEXT to switch the pump ON and continue warm-up cycle.
If the monitor fails the flow fault test, the screen shown in
Note: The monitor logs a successful Flow Fault test or if SKIP was selected.

3.5. Time and Date

The time and date are displayed as shown in
they are used with datalogging.

3.6. Calibration Due

Calibration of the monitor is important.
By default, CAL DUE date is set to 365 days, but can be set to from 1 to 400 days.
The CAL DUE date is updated when the monitor is successfully calibrated.
The monitor has five options to alert the user when calibration is due or is overdue:
1. Cal Due disabled - no date is displayed. The monitor does not require a valid calibration to
operate.
2. Cal Due enabled - if overdue, monitor is switched OFF.
3. Cal Due enabled - if overdue, message is displayed as shown in
Date, but after five seconds, the monitor warm-up continues automatically.
14
Figure 6: Successful Test
Figure 7: Time and
Figure 7: Time and Date
Figure 8: Calibration Due Date
Figure 5: Flow Fault Test
remains displayed.
Date. Verify they are correct, as
Figure 8: Calibration Due
67816
Revision D