Sample / Flow Fault - Teledyne Everywhereyoulook GT-FIRE User Manual

Portable gas detector
Table of Contents

Advertisement

• Switch monitor OFF and ON again.
• If the fault persists, return the monitor for service.
2. 'ZERO FAULT' and a wrench symbol, alternating with a gas reading, as illustrated in
Zero Fault
(LEL).
• Leave monitor ON for 30 to 60 minutes.
• Switch it OFF and ON.
• If the fault persists, return the monitor for service.

5.5.4. Sample / Flow Fault

If a sample fault occurs, 'FLOW FAULT' alternates with the mode name, as shown in
Fault
(CSM).
• Pump switches off (except Confined Space mode)
• Pump symbol disappears
• Fault LED tuns ON.
Check sample line, filter or probe for blockage. Clear blockage then restart the pump by pressing
the FLOW ACK
button.
67816
Revision D
Figure 28: Zero Fault
Figure 29: Flow Fault (CSM)
GT-FIRE
PORTABLE GAS DETECTOR
Figure 27:
Figure 29: Flow
27