Operating the System
Wallstand Examination Procedures

4.10.2 NoWait Configuration

WARNING!
The Wallstand will immediately start tracking the height of the detector.
At the installation of the System, it is possible to select that the System shall not wait for the
user to move the Wallstand.
The flow will then be as follow;
Select a Wall Flexible mode examination.
• The SID will be displayed.
Activate the mode by pressing the servo button.
• All movements are available.
• The OTC will automatically move to its programmed position.
• The OTC will move down and align with the Wallstand detector.
• The OTC will follow the Wallstand detector to stay aligned.
• The OTC reaches the correct height and stops.
• The SID value is changed to the new distance (if selected distance towards the Wallstand
is changed).
4.10.2.1 Exposure
Exposure is possible when the OTC stands still, and the servo mode indication light is fixed.

4.10.3 Stitching Wall Stand Mode

Note!
Stitching Wallstand Mode is not available according to the described procedure with CR
systems.
WARNING!
The Wallstand detector holder will move during stitching. This may cause danger for
the patient.
Select a Stitching Wallstand mode examination.
The System display will display the following.
OMNERA®
1000-095-071-EN Rev. 1.1
Page 128
Operator´s Manual
2020-09-14