4.10.1.3 Basic Flow
Select a Wall Flexible mode examination.
The System display will display the following.
• The SID will be displayed.
Activate the mode by pressing the servo button.
• All movements are available.
• The OTC will automatically move to its programmed position.
• The OTC will either go direct to an aligned position to the detector or wait in the transporta-
tion height until detector is moved (wait or no wait selected).
• The mode will stay activated. The OTC beeps once and the servo mode indication light will
flash.
Adjust the height of the wall stand.
Move the detector holder at the wall stand.
• The OTC will move down and align with the wall stand.
• The OTC will follow the wall stand to stay aligned.
• The OTC turns on the collimator light when the correct height is reached.
Adjust the position.
• The mode will stay activated.
• The SID value is changed to the new distance (if selected distance towards the wall stand
is changed).
Operator´s Manual
2020-09-14
Operating the System
Wallstand Examination Procedures
Fig. 4-31 Wall Flexible mode display
1000-095-071-EN Rev. 1.1
Page 127