Canon Omnera 500A Operator's Manual page 60

Hide thumbs Also See for Omnera 500A:
Table of Contents

Advertisement

Safety
Electromagnetic compatibility (EMC)
Guidance and manufacturer's declaration - electromagnetic immunity
Immunity test
IEC 60601 test
level
Note!
U
is the AC mains voltage prior to application of the test level.
T
3 Vrms
Conducted RF
IEC 61000-4-6
150 kHz to 80
MHz
6 Vrms (ISM and
amateur radio
bands)
Radiated RF IEC
3 V/m
61000-4-3
10 V/m
80 MHz to 2.7
GHz
Proximity field
9 V/m to 28 V/m
from wireless
15 specific
transmitters
frequencies
61000-4-3
NOTE 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
OMNERA®
Electromagnetic environment -
Compliance
guidance
level
Portable and mobile RF communications
equipment should be used no closer to any
part of the system, including cables, than
the recommended separation distance,
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance;
3 Vrms
d = 1.2 √p
150 kHz to 80
MHz
6 Vrms (ISM and
amateur radio
bands)
3 V/m
d = 1.2 √p 80 MHz to 800 MHz
10 V/m
d = 2.3 √p 800 MHz to 2.7 GHz
80 MHz to 2.7
where p is the maximum output power rat-
GHz
ing of the transmitter in watts (W) accord-
ing to the transmitter manufacturer and d is
the recommended separation distance in
metres (m).
9 V/m to 28 V/m
15 specific
frequencies
Interference may occur in the vicinity of
equipment marked with the following sym-
bol:
1000-095-071-EN Rev. 1.1
Page 52
Operator´s Manual
2020-09-14

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents