User Interface
Table Control elements
3.7.10.1 Synchronization Function
In Auto position mode and in Table flexible mode, the detector holder may be moved in X-di-
rection using the motor (via the table handle or manually by activating the green button at the
detector holder). There is no synchronization between the tube and the table detector.
In Film tracking and Pendulum mode the detector holder may be moved manually in X-direc-
tion. If this is done, the servo button will be deactivated and exposure is no longer possible.
To return to an activated servo button, the detector holder shall be moved to the correct posi-
tion. The correct position will be indicated by the lightning of the green detector holder
button.
Note!
It is the user's responsibility to verify that the detector is in position at exposure.
OMNERA®
1000-095-071-EN Rev. 1.1
Page 96
Operator´s Manual
2020-09-14