4.4.8 Basic exposure error handling

Exposure not possible
The small detector is se-
lected (Green)
Table examination
Table or wall stand
Table examination
Wall stand examination
Operator´s Manual
2020-09-14
Operating the System
Check
Is the small detector in the
docking station?
If the table is equipped with
a wireless detector and
charging the detector in the
holder - check if the connec-
tor is correctly connected to
the detector.
Is the indication light lit (yel-
low LED in the control room
or green light on the OTC)?
Is the detector in the table
detector holder?
Is the detector in the wall
stand detector holder?
1000-095-071-EN Rev. 1.1
Page 113
Perform examination
Measure
Remove the small detector
from the docking station.
Connect the connector cor-
rectly to the wireless
detector.
Position the system correct
by pressing the servo button
on the remote control, the
operators console or on the
OTC.
or
Change from Auto to Man-
ual on the OTC display if the
patient/light field is in posi-
tion (and you don´t want to
reposition the OTC).
Place the detector in the ta-
ble detector holder, make
sure to connect the connec-
tor correctly.
Place the detector in the
wall stand detector holder,
make sure to connect the
connector correctly.