Automatic Collimator - Canon Omnera 500A Operator's Manual

Hide thumbs Also See for Omnera 500A:
Table of Contents

Advertisement

User Interface
Automatic Collimator Control

3.2.1 Automatic collimator

1. Select Automatic or Manual mode of the collimator.
Note!
If there is no new examination and the System is in a Manual mode, the collimator
ignores the instructions of sent collimator area from the Imaging system. This can be use-
ful if the collimator area is intended to be kept for the next examination.
2. Collimator light/x-ray field height x width
3. Collimator filter selection
3.2.1.1 Collimator mode
1
Fig. 3-3 Collimator mode
When Automatic mode is selected the predefined values of the collimator light / x-ray field
size and the filter selection will be set automatically when the anatomical protocol is selected.
Both the light / x-ray field size and the filter can be changed when needed. In automatic mode
the maximum light / x-ray field size is restricted to the maximum active area of the imaging
unit.
When manual mode is selected the size of the collimator light / x-ray field size can be ad-
justed outside of the maximum active area of the imaging unit. When a new anatomical pro-
tocol is selected (for the same patient) the collimator light / x-ray field size or the filtration is
not changed even if size and filtration is defined different in the protocol.
OMNERA®
DoB 1977-03-06
ID 987-65-4320
Jane Doe
Age 36 Sex F
Acc No 987-65-4320
Abdomen suspine
60
60
110
cm
50
250
20
8
mAs
kV
mA
ms
x
43.0
43.0
cm
cm
1
2
In automatic mode, the collimator light
height and width is set from the anatomical
program. The collimator light size can still
be changed manually. In manual mode the
collimator light height and width is not set
from the anatomical program.
1. Automatic mode
2. Manual mode
2
1000-095-071-EN Rev. 1.1
Page 58
i
1
AEC
Density
1.0
Al
0.1
Cu
3
Operator´s Manual
2020-09-14