Immunity
IEC 60601
test
test level
Conducted
3 V
dis-turbances
0.15 MHz – 80
induced by RF
MHz, 6 V in
fields
ISM bands be-
IEC 61000-4-
tween 0.15
6
MHz and 80
MHz, 80 % AM
at 1 kHz
3 V/m, 80 MHz
– 2,7 GHz,
80 % AM at 1
kHz
Radiated RF
EM fields
IEC 61000-4-
3
See the RF
wireless
communication
equipment
table in
"Recommende
d minimum
separation
Proximity
distances"
fields from RF
wireless
communicatio
n equipment
IEC 61000-4-
3
1. Use of accessories and cables other than those specified or provided by the
manufacturer of E-connect could result in increased electromagnetic emissions
or decreased electromagnetic immunity of E-connect and result in improper
operation.
Compliance level
3 V
3V/m
Complies
37
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment
should be usedno closer to any
part of the E-connect , including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended minimum
separation distances
See the RF wireless
communication equipment table in
"Recommended minimum
separation distances"