Packaging
8
Sterilization
Black oil
Before autoclaving, the contra angle must be
●
lubricated.
Attaching the spray nozzle to oil can and contra
●
angle, press the oil can button more than 3
seconds, till all the black oil flow out from the
head of the contra angle.
Pack each component in a separate steam-
sterilization pouch.
Check the validity period of pouch given by the
●
manufacturer to determine the shelf life.
Use pouches which resist to a temperature up
●
to 141℃(286°F) and in accordance with EN
ISO 11607.
Steam sterilization at 134℃, at least 6 minutes.
Minimum drying time after sterilization: 10 minutes.
Use only approved autoclave devices according
●
to EN 13060 or EN 285.
Use a validated sterilization procedure
●
according to ISO 17665.
Respect the maintenance procedure of the
●
autoclave device given by the manufacturer.
Use only this recommended sterilization
●
procedure.
Control the efficiency (packaging integrity, no
●
humidity, color change of sterilization indicators,
physicochemical integrators, digital records of
cycles parameters).
The sterilization procedure must comply with
●
ISO 17665.
Wait for cooling before touching.
●
31