AT510x DC Resistance Meter
16
Range Hold is help increasing the measurement rate..
In the "automatic range" status, the meter should be calibration (zeroing) when it can't
Tip:
choose the fit Range.
Full calibration (Zeroing) in the "Calibration".
In the range hold status ("AUTO" indication is OFF), the meter will choose the fit range by
the upper limit value.
3.4.4
Speed
Full Sampling Speed in the Appendix A.
3.4.5
Calibration (Short Zeroing)
Figure 3‐7
Right Way
Figure 3‐8
Wrong Way
3.4.6
Relative Value for Temperature Compensation
3.4.7
‐
% (The function is invalid in the version.)
3.4.8
Keypad Locked
9
1M
10
10M
Press Range < > keys to choose the fit range.
Press Rate key to choose the following speed :
Fast:
Fast Sampling
Medium: Medium Sampling
Slow:
Slow Sampling
1 Press Clear key to the Clearing status and make the test clips short-circuit like the
.
following way before zeroing.
2 Press Enter key to Calibration.The meter zeroing all ranges in the automatic range or
.
the present range in the hold range.
Data will be saved in the nonvolatile memory when zeroing is over.
3 Press Esc or when zeroing is over, return to the measurement status.
.
AT510pro, AT510 (Standard), AT510se (Optional), AT510L, AT510M (invalid)
Press
Shift
Relative key to Turn ON/OFF the relative function. The REL indicator is OFF;
the value is the real measurement value.
Press Shift Key Lock key to lock/unlocked the keypad.
The Key Lock key is only available and the other keys are in vain when the keypad is
locked.
Need help?
Do you have a question about the Anbai AT510PRO and is the answer not in the manual?