8 P
/PCM O
ROBE
PERATION
Q
C
UICK
APTURE
KEY
NOTE:
56
Fig. 22
FBPT series tips for the P5000i
1
Inspection tip
2
Barrel assembly
P5000i connection
The Probe application requires a P5000i Digital Probe in order to
be fully functional (see the list of all accessories in
Microscope" on page
1.
Plug your P5000i into an USB port.
2.
Connect the P5000i with the fiber being inspected.
[C]
3.
Press the
key, then tap the
arrows keys.
– or –
[H]
Press the
key, to toggle between the measurement view
and the inspection view.
The P5000i will display the live view.
QuickCapture
In order to support different workflows the functionality of the
QuickCapture™ key is configurable. Pressing the key will either
freeze the live image or start a test:
Test
Pressing the key will automatically freeze the live
image and perform a test.
Freeze
Pressing the key will automatically freeze the live
image.
Camera mode
Set the QuickCapture™ key on the Digital Probe to get a live
picture of the result.
Use the Focus Control key to adjust the focus of the image.
The Digital Probe is connected to the instrument.
I
89).
[Probe]
key
"Digital Probe
button or use the
OLP-87/OLP-87P