Test Movements (Advanced)
The
Test
Movements
window
provides a way to test each available
movement in the prosthesis, without
the need for calibration. By clicking the
button associated with a movement,
it will activate the selected movement
for 4 seconds while preventing other
buttons from being pressed.
WARNING: Avoid contact with the
prosthesis components while in Test
Movements.
Velocity Settings (Advanced)
Velocity changes the signal voltage
output to the prosthesis to increase or
decrease the speed of the hand, wrist,
or elbow.
Select a movement and choose a
value between 10 and 100 in the
popup menu.
Open/Close Mode
(Located in Settings tab)
When
this
mode
is
active,
calibrated grip movements except
Open will close the hand in the grip
last used before the "Open/Close
Mode On" toggle was pressed. The
only exception is the Open movement,
which will open the hand in that grip.
In order to change grips, you can use
the buttons on the back of the hand or
an App that comes with the hand.
70
all
Proportional Control (Advanced)
The
Proportional
control
switch
turns proportional control on or off.
Proportional control allows the patient
to change the velocity of any given
movement while the patient controls
it, by changing the force of his/her
muscle contractions.
This means,
that if for example the patient mimics
closing his/her phantom hand more
forcefully, the prosthetic hand will close
faster and vice versa. The proportional
control functionality is off by default. It
is white when off and red when on.
Enable Advanced Settings
You can hide certain settings within
the Settings tab if you do not wish for a
patient to adjust them. Upon selecting
"Enable Advanced Settings", you will
be prompted with a password. After
you enter the password (the same
password in the previous section
"Device Configuration" on Page 63) it
will allow you show or hide electrode
settings, velocity control, proportional
control, and Test Movements.
In the event that the Sense controller is
not functioning correctly, you can delete
the Database and Device Configuration.
Please contact IBT before proceeding
with either of these actions as it can
erase all date from the prosthesis.