testing sense

5
Please test the Sense system before fitting the patient with the prosthesis.
Connect all electrodes, prosthesis components, and FlexCells. Power on
Sense. It will vibrate twice and then once more after initialization. Connect to
the Sense software and view the signals by tapping each electrode contact
with your fingertip. You should observe peaks generated every time you tap
an electrode contact.
Navigate to the Calibration Setup page and click on the Test button (see
software descriptions below). When pressing buttons associated with
certain movement, observe if the terminal device, wrist and elbow respond
as expected (i.e. wrist supinates after button is pressed). Reverse output
cables if the movements are backwards.
If the Sense does not vibrate, peaks are not seen, or the components are
not reacting, confirm that all cables are securely connected to the Sense
controller and all the other prosthetic components.
Note: The patient must first use the device in the presence of a
trained practitioner.
36

sense GUI

6

CONNECTING SENSE TO SOFTWARE

Connecting the Sense System to the Sense GUI
The Sense system should already be paired to the Microsoft Surface. Open
the Sense GUI.
A Login screen will appear. Select the "Practitioner" ID and type "Sense2017"
(no quotations) as the password. Press Enter.
Note: Practitioner's accounts will have several options available that will not
be present for Client accounts. Further on these options will be marked with
"Practitioner Only".
Bluetooth Connection
The User Interface will automatically begin searching for Bluetooth devices.
A window will appear listing all available devices. Select the name of the
Sense system provided with your order from the list (IBTCoreXXXXXX) and
click OK.
Note: Sense cannot connect to Catalyst or the IBT App while an attached
TASKA Hand is connected to TASKA's HandCal Clinician software. Please
close HandCal Clinician before connecting to Sense.