The color indicators for selecting movements, described above, match that
for selecting positions.
Note: If the movements selected change, the color indicators on the position
buttons may change as well.
Calibration Options
Under the column labeled "Calibration" there are three options for modifying
the classifier:
Calibrate
When this option is selected, you will proceed directly to the calibrate screen.
All of the movements and positions selected will be recalibrated. The local
data will be replaced with new data for all movements.
Custom Calibration
When this option is selected, another small window will appear listing all
movement and positions that are selected on the Setup page. Select only
the movements and/or positions that you would like to recalibrate. All
movements and positions listed on this pop up will be a part of the final
classifier. However, only the ones that are selected will be recalibrated to
have new data replace the old data.
46
Note: If a movement or positions is outlined in black or red on the Setup
page, you are not able to deselect it.
Update Classifier
If a movement or position is deselected OR if a movement or position is
selected that is filled in red, select Update Classifier to change the movements/
positions selected without recalibrating.
Advanced Movement Settings (Practitioner Only)
The options in this window allow you to fine tune the Sense control. Each
movement calibrated has its own specific settings for Sensitivity and Velocity.
You can also turn proportional control on or off from this page.
Sensitivity
Sensitivity improves the smoothness of control over a movement. Increasing
the sensitivity will make the movement more responsive, but has the potential
to cause more accidental activation. Decreasing the sensitivity makes the
movement more stable, but may cause a small delay before a response by
the prosthesis.