INSTALLING SENSE INTO THE PROSTHESIS

Sense has the following electrical output characteristics:
EMG Signal (5V Envelope)
Sense is a pass-through for the battery, therefore output voltage
matches input battery voltage (5-10V)
Motor Driver Output (matches input battery voltage)
All terminal devices that accept the aforementioned electrical outputs should
be compatible with Sense. For a list of terminal devices that have been
tested with the Sense system please refer to page 10 of the user manual and
contact IBT for an up-to-date list of all compatible devices that have been
tested.
Now that the electrodes and battery are connected, insert Sense through the
proximal end of the prosthesis.
WARNING: For all setups below, ensure the battery is OFF before
plugging in any cables. Do not attempt to connect the Sense output
cables in a way that is not specified. This could cause permanent
damage to the prosthesis and Sense.
24
Connecting to the terminal device, wrist, and elbow
The following are the connections to some of the commonly used prosthetic
devices. For connections to all other devices, contact IBT.
* Data cable availability is device-dependent