9640A
Instruction Manual
Modulation (Optional)
Use a high performance measuring receiver to make the following measurements.
+10dBm
+14dBm
+13dBm
+13dBm
5-66
Level
Frequency
30 MHz
Level
Frequency
125 MHz
1 GHz
Level
Frequency
1 MHz
1 GHz
Frequency Modulation Deviation Test
Level
Frequency
125 MHz
1 GHz
Amplitude Modulation Rate Test
Modulation
Depth
Rate
1 kHz
50 %
220 kHz
50 %
Amplitude Modulation Depth Test
Modulation
Depth
Rate
1 kHz
80 %
100 kHz
80 %
1 kHz
80 %
100 kHz
80 %
Frequency Modulation Rate Test
Modulation
Deviation
Rate
1 kHz
300 kHz
300 kHz
1 MHz
Modulation
Deviation
Rate
1 kHz
100 kHz
100 kHz
100 kHz
1 kHz
300 kHz
200 kHz
300 kHz
1 kHz
1 MHz
300 kHz
1 MHz
Actual
Tolerance
± 0.1 Hz
± 10.0 Hz
Actual
Tolerance
± 2.5 %
± 3.1 %
± 3.1 %
± 3.1 %
Actual
Tolerance
± 10 Hz
± 10 Hz
Actual
Tolerance
± 3.0 kHz
± 3.0 kHz
± 9.0 kHz
± 9.0 kHz
± 30 kHz
± 30 kHz