Output Subsystem - Fluke 9640A Series Instruction Manual

Only to older units with firmware up to lss 2.x
Table of Contents

Advertisement

9640A
Instruction Manual

OUTPut Subsystem

:OUTPut
:OUTP[:STATe][?]
:OUTP:ROSCillator[:STATe][?]
:OUTP:ROSCillator:FREQuency[?]
:OUTP:FITTed?
4C-6
Keyword
Parameter
*RST
Form
Condition
<bool>
OFF
<bool>
Unchanged
<NRf>
Unchanged
<SPD>,<SPD>
n/a
Notes
Command long form
The STATe command
controls whether the output
terminals are open or
closed. When the state is
OFF, the terminals are at
maximum isolation from
the signal.
The STATe command
controls whether the
reference frequency is
output on the rear panel
BNC
Selects the output
frequency frequency on the
back panel, in Hz
Query only command that
returns two strings, the first
is the head model type and
the second is the serial
number. If no head is fitted
then the query will return
"NONE","NONE"

Advertisement

Table of Contents
loading

This manual is also suitable for:

9640a-lpn9640a-std

Table of Contents