9640A
Instruction Manual
Additional notes regarding good practice when sourcing and measuring high- and low-
level signals are given at the end of this chapter.
Use the following procedure to connect a Leveling Head to a UUT:
1. Read and observe all of the preceding Cautions and Warnings.
2. Remove the plastic connector protection caps form the cable-end connectors and save
3. Connect the N-connector on the Leveling Head to the input of the UUT.
4. Torque the N-connector to 1.00 Nm (9 in-lb) using an N-connector torque wrench.
Using the Save/Recall Function
The Save/Recall function provides the user with a way to save and recall up to 10 groups
of settings associated with the instrument setup and/or the output signal.
Each memory group has a default name, SLOT-1 through SLOT-10, and is accessible
from the Instrument Setup screen. From this screen, the user can do the following:
•
•
•
•
Accessing the Memory Screen
To access the Save/Recall screen press . When the screen initially comes up it is
ready to perform save/recall operations on the selected (SLOT-1) memory. These
operations include Rename, Delete, Save Instrument, Save Mode, and Recall Settings. A
description of each operation follows:
3-34
them for future use.
The torque wrench is available as an accessory; see Chapter 1, Options and
Accessories.
Save the present Instrument or output signal settings to a selected slot
Recall previously saved instrument or output signal settings from a selected slot
Rename a selected memory slot to something more meaningful
Delete all setting information from a selected memory slot
Save/Recall Screen
ead11f.bmp