Setup ® Advanced setup ® Analog output 1...4 ® Fail safe type
Defines the output behavior of the AO block in the case of an error.
• Fail safe value
The output assumes the value defined in Fail safe value.
• Fallback value
The output maintains the value it had at the moment the error occurred.
• Off
The output assumes a value according to the current state of the block. The error is ignored.
Fallback value
Setup ® Advanced setup ® Analog output 1...4 ® Fail safe value
Only visible for Fail safe type = Fail safe value
Defines the value the output assumes in the case of an error.
Any IEEE floating point number:
-3,4 10
38
... + 3,4 10
38
0
Setup ® Advanced setup ® Analog output 1...4 ® Out value
Diagnostics ® Measured value ® Analog output 1...4 ® Out value
Displays the current output value of the block.
Setup ® Advanced setup ® Analog output 1...4 ® Out status
Diagnostics ® Measured value ® Analog output 1...4 ® Out status
Displays the current status of the output value of the block.