List Of Parameters - Culligan MP97 Manual

Table of Contents

Advertisement

Available languages

Available languages

SERIAL LINE PROGRAMMING
The procedure for programming the 22 parameters (all those available) which can be modified from the RS232C
(serial line) is as follows. Please note that this programming can be carried out by STEIEL only (the serial line is not
available on the instrument) :
1.
Connect the programming keyboard TP92 or a PC furnished with serial communication software to the
Cannon 9-pin female connector. We can supply simple communication software on request.
2.
Digit [ ED (CR) ].
3.
<<Pro>> appears on the µP 97 display while the description and value of P01 (first parameter) appears on
the programming keyboard.
4.
If the value is the one desired, press [ (CR) ] to pass on to the next one. If the value is to be changed, digit
the new numerical value and press [ (CR) ] to confirm.
5.
If the value is included in the range it will be accepted and the µP 97 will propose the next parameter.
Otherwise, an error measure will appear and the same parameter will reappear.
6.
And so on up to parameter 22 (the last).
NOTES:
1)
To exit from programming enter [ E ] or scan all parameters to the last.
2)
To see a preceding parameter again, digit [ S ].
3)
To see a specific parameter :
a) digit [ X ]
b) the µP 97 responds with <<PARAMETER ?>>
c) digit the number of the desired parameter.
4)
All serial commands require capital letters.
5)
Values must be entered without commas or decimal points.
6)
The cancel or delete command is not activated; if an error is made entering data, the value must be called
up again and rewritten.

LIST OF PARAMETERS

We advise filling in the last column with the values set for the application in question.
Parameter Description
P01
Type of output relay 1
P02
Min. threshold relay 1
P03
Max. threshold relay 1
P04
Energization delay relay 1
P05
Deenergization delay relay 1
P06
Type of output relay 2
P07
Min. threshold relay 2
P08
Max. threshold relay 2
P09
Energization delay relay 2
P10
Deenergization delay relay 2
P11
Type 1° current output
P12
S.S. 1° current output
P13
F.S. 1° current output
P14
Type 2° current output
P15
S.S. 2° current output
P16
F.S. 2° current output
P17
Initial value 1° current output
P18
Initial value 2° current output
P19
Conversion coefficient
P20
Measurement start scale
P21
Measurement full scale
P22
Measurement point (comma) position
NOTES
Unless otherwise requested, parameters 20 to 22 can only be modified from the keyboard since they can
compromise the operation of the instrument if they are set wrongly.
M003-23 - Rev. 03 - 05/2009
Min.
Max.
value
value
0
-5000
-5000
0
0
0
-5000
-5000
0
0
0
-5000
-5000
0
-5000
-5000
0
0
0
-5000
-5000
0
Microprocessore µP97
Set
value
3
25000
25000
28800
28800
3
25000
25000
28800
28800
1
25000
25000
1
25000
25000
2000
2000
400
25000
25000
3
10