Maintenance; Maintenance Schedule; Disinfecting And Cleaning The Table Top; Disinfecting And Cleaning The Table-Base Cover - Ferno 101-H User Manual

Operating table
Table of Contents

Advertisement

Model 101-H Operating Table

6.1 Maintenance schedule

The table requires regular maintenance. Set up and follow a
maintenance schedule. The table at right represents minimum
intervals for maintenance.
Keep maintenance records. A sample maintenance record sheet
is provided on page 17.
When using maintenance products, follow the manufacturers'
directions and read the manufacturers' material safety data
sheets. You can purchase a recommended disinfectant from your
Ferno distributor or Ferno Customer Relations (page 2).
6.2 Disinfecting and Cleaning
the Table Top
Disinfectinge: Wipe all surfaces with disinfectant. Follow the
disinfectant manufacturer's instructions for application method
and contact time. Ferno recommends you inspect the table top
for damage as you disinfect it.
Cleaninge: Hand clean all surfaces of the stainless steel table
top with warm water and a mild detergent. Rinse well with clear
water. Dry the table top with a towel or allow it to air-dry.
6.3 Disinfecting and Cleaning
the Table-base Cover
Disinfectinge: Disinfect the table-base cover with a disinfectant
suitable for use on ABS plastic.
Cleaninge: Hand wash the ABS plastic table-base cover with a
mild detergent and warm water. Rinse well with clear water and
dry with a towel.

6.4 lubrication

The table is manufactured with parts that do
not require lubrication.

6.5 inspecting the Table

Have your service's equipment maintenance personnel inspect
the table regularly. Follow the checklist at right and operate the
table through all its functions as described in this manual.
If inspection shows damage or excessive wear, remove the
table from service until repair is made. See Parts and Service,
page 14 and page 15.
© Ferno-Washington, Inc 234-1994-03 October 2010

6 - MainTenanCe

improper maintenance can cause injury. Maintain
the table only as described in this manual.
Disinfecting (this page)
Cleaning (this page)
inspecting (this page)
Lubricating parts that should not be lubricated allows
dirt and foreign particles to collect on those parts,
resulting in damage. Do not lubricate any part of the
table top, wheels, jacks, or base.
Do Not
Lubricate
Warning
Minimum
Maintenance intervals
important
Do not use cleaning solvents on the table-base cover.
Cleaning solvents can damage the ABS plastic.
Do not use disinfectants and cleaners containing
bleach, phenolics, or iodines on the table top, base, or
base cover. These chemicals can damage the table.
Do not use abrasive applicators or cleaners on
the table top, base, or base cover. Abrasives in
any form can damage the table.
important
insPeCTiOn CHeCklisT
Are all components present?
Are the table top, table base, jacks and wheels free of
excessive wear?
Are all screws, nuts, bolts, rivets, and roll pins
securely in place?
Do all moving parts operate smoothly and properly?
Do the pedals operate properly?
Does the table raise, lower, and tilt properly?
Does the wheel-locking system function properly?
Does the table roll smoothly?
Maintenance