5.1 Maintenance Schedule

The IV pole requires regular maintenance. Set up and follow a
maintenance schedule. The table at the right represents minimum
intervals for maintenance.
Keep maintenance records. A sample maintenance record sheet is
provided at the back of this manual. See "Maintenance Records, " on
page 14.
When using maintenance products, follow the manufacturers'
directions and read the manufacturers' material safety data sheets.
The IV pole is constructed using parts and materials that require
no lubrication. Only trained service personnel are permitted to
disassemble and service the IV pole.

5.2 Disinfecting and Cleaning the IV Pole

DISINFECTING
Wipe all surfaces with disinfectant. Follow the disinfectant
manufacturer's instructions for application method and contact
time. Ferno recommends you inspect the IV pole for damage as you
disinfect it.
CLEANING
1.
Hand clean all surfaces of the IV pole with warm water and a
mild detergent.
2.
Rinse with warm, clear water. Dry the IV pole with a towel or
allow it to air-dry.
5.3 Disinfecting and Cleaning the IV Bag
Strap
Remove the IV bag strap following the steps in "Attaching the IV Bag
Strap, " page 9. Spray all parts of the IV bag strap with disinfectant.
Follow the disinfectant manufacturer's directions for use. Place the
strap in a mesh bag (not supplied) and machine wash in warm water
with a mild detergent. Hang the IV bag strap to dry (do not place the
IV bag strap in a dryer).

5.4 Inspecting the IV pole

Have your service's equipment maintenance personnel follow the
instructions "Using the IV Pole, " pages 7-10, and follow the inspection
checklist at the right.
The inspection limits listed at the right are maximum allowable
damage and are not to be exceeded. Item function takes precedence
over allowable damage. If the unit has greater damage than what is
listed, or does not function correctly, then it must be removed from
service and replaced.
© Ferno-Washington, Inc. / 00271-V3
Maintenance

5 -MAINTENANCE

Improper maintenance can cause injury. Maintain the IV
pole only as described in this manual.
Disinfecting
Cleaning
Inspecting
Disinfectants and cleaners containing bleach, phenolics or
iodines can damage the IV pole and straps. Disinfect and
clean the IV pole and straps only with products that do not
contain these chemicals.
Water under high pressure can penetrate joints, flush away
lubricant and cause corrosion. Do not use a high-pressure
washer or steam, to clean the IV pole.
Cracks
Dents
Scratches and nicks
WARNING
Minimum
Maintenance
Each Use
Important
Inspection Checklist
Are all components present?
Are the knurled pins in the clamps firmly in place?
Do all moving parts operate smoothly and properly?
Does the pole operate properly without interfering with
the cot or iN∫X operation?
Is the IV bag strap in good condition with no cuts or
frayed edges?
Type of Damage
As Needed
Monthly
Permissible Damage
None
Up to 1 mm
OK
11