Agilent Technologies 5975 Series Operating Manual page 79

Msd for openlab cds
Hide thumbs Also See for 5975 Series:
Table of Contents

Advertisement

4
Operating in EI Mode
To view the firmware version of the MSD from
the GC control panel
1 Press MS/Aux Det to bring up the 5977B MSD menu.
2 Press the down arrow to scroll to Firmware.
To view the serial number of the MSD from the
GC control panel
1 Press MS/Aux Det to bring up the 5977B MSD menu.
2 Press the down arrow to scroll to Serial#.
To configure the network settings for the MSD
from the GC control panel
1 Press Config, and then press MS/Aux Det to bring up the CONFIGURE MS
2 To configure the IP: parameter, use the GC keypad to enter the new IP
3 Wait for the GC to display the new IP address. Reboot the MSD or proceed to
4 Press the down arrow to scroll to GW: and use the GC keypad to enter the
5 Press the down arrow to scroll to SW: and use the GC keypad to enter the
6 Reboot the MSD. (See below)
To reboot the MSD from the GC control panel
1 Press Config, and then press MS/Aux Det to bring up the CONFIGURE MS
2 Press the down arrow to scroll to Request MSD Reboot?.
3 Press On/Yes to reboot the MSD and wait for the MSD to complete this cycle
5975/5977 Series MSD Operating Manual
DETECTOR menu.
address for the MSD, then press Enter to complete the entry.
the gateway address with the down arrow button
new gateway address for the LAN and press Enter to complete the entry.
new subnet mask for the LAN and press Enter to complete the entry.
DETECTOR menu.
before trying to access it.
79