General Maintenance
3 Collect the following parts to be cleaned for an EI XTR source: (See
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
If insulators are dirty, clean them with a cotton swab dampened with
C AU T I O N
reagent-grade methanol. If that does not clean the insulators, replace them. Do
not abrasively or ultrasonically clean the insulators.
148
on page 149.)
Repeller
Repeller block insert
Source body
Extractor lens
Ion focus lens
Entrance lens
5975/5977 Series MSD Operating Manual
Figure 46