General Maintenance
Some parts can be damaged by electrostatic discharge
The wires, contacts, and cables connected to the analyzer components can carry
electrostatic discharges (ESD) to the electronics boards to which they are
connected. This is especially true of the mass filter (quadrupole) contact wires
which can conduct ESD to sensitive components on the side board. ESD damage
may not cause immediate failure but will gradually degrade performance and
stability. (See
Electrostatic discharges to analyzer components are conducted to the side
C AU T I O N
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap (See
other antistatic precautions before you open the analyzer chamber.
Some analyzer parts should not be disturbed
The mass filter (quadrupole) requires no periodic maintenance. In general, the
mass filter should never be disturbed. In the event of extreme contamination, it
can be cleaned, but such cleaning should only be done by a trained Agilent
Technologies service representative. The HED ceramic insulator must never be
touched.
Incorrect handling or cleaning of the mass filter can damage it and have a
C AU T I O N
serious, negative effect on instrument performance. Do not touch the HED
ceramic insulator.
More information is available
If you need more information about the locations or functions of analyzer
components, refer to the Agilent 5977B Series MSD Troubleshooting and
Maintenance Manual.
Many procedures in this chapter are illustrated with video clips.
118
"Electrostatic discharge"
"Electrostatic discharge"
on page 115 for more information.)
on page 115.) and take
5975/5977 Series MSD Operating Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5977 series5977b

Table of Contents