Agilent Technologies 5975 Series Operating Manual page 158

Msd for openlab cds
Hide thumbs Also See for 5975 Series:
Table of Contents

Advertisement

5
General Maintenance
5 Secure the new filaments with the gold plated screw and washer. (See
6 After installing the filament, verify that it is not grounded to the source body.
7 Install the EI source. (See
8 Close the analyzer chamber. (See
9 Pumpdown the MSD. (See
10 Autotune the MSD. (See
11 In the Manual Tune dialog, the Filament parameter allows you to enter 1 or 2
12 Autotune the MSD again.
13 Enter the filament number that gave the best results.
14 Select Save Tune Parameters from the File menu.
158
Figure 49
on page 157.)
page 159.)
page 162.)
for the filament number. Whichever number was present during the previous
autotune enter the other filament number.
If you decide to use the first filament number, run Autotune again to make
sure the tune parameters are compatible with the filament.
"To Install the EI XTR, SS, or Inert Source"
"To Close the Analyzer Chamber"
"To Pump Down the MS"
"To Run an Autotune"
5975/5977 Series MSD Operating Manual