Intuvo 9000 GC Column Maintenance
The 9000 GC/MS Tail is heated by an electric cartridge heater. The heater is
powered and controlled by a heated zone of the 9000 GC. The 9000 GC/MS Tail
temperature can be set from the Agilent MassHunter GC/MS Acquisition
software or from the GC. A sensor (thermocouple) in the interface monitors the
temperature.
The transfer line tip seal assembly is required when using the EI XTR source or
the HES.
The 9000 GC/MS Tail should be operated in the 250  to 350 C range. Subject to
that restriction, the transfer line temperature should be slightly higher than the
column temperature, but never higher than the maximum column temperature.